Instrument image. The configuration and appearance of the supplied equipment are agreed separately.

ID: 00591

Water degassing and temperature conditioning unit BT STTQS-1

Prepares degassed, temperature-conditioned water for hydraulic geosynthetics testing.

Materials: Not assigned

Capacity50 L
Temperature range
15-105 °C
Stated temperature calibration accuracy
±0.1 °C

Application

The unit supplies prepared water to connected test equipment. Removing dissolved gases supports consistent hydraulic test conditions, while heating and cooling allow water temperature to be matched to the procedure.

It is a laboratory support unit rather than a standalone permeability instrument. Check compatibility against the required flow, temperature and connections.

Specify the connected instruments and required water flow and temperature. Agree the circulation arrangement and heat rejection conditions, and confirm the accuracy designations and cooling rating for the supplied configuration.

Specifications

ParameterValue
Capacity50 L
Temperature range15-105 °C
Stated temperature calibration accuracy±0.1 °C
Temperature accuracy, stated designation+2 °C
Stated cooling rating13000 W
Compressor power1.5 HP
Circulation pump power100 W
Circulation pump flow45 L/min
External circulation pump100 W, adjustable flow
Approximate heating time to boiling30 min from room temperature
Approximate cooling time90 min from boiling to 20 °C
CirculationInternal and external
Temperature conditioningHeating and cooling
Stated cooling arrangementAir conditioner outdoor unit
ControlPID, LCD touchscreen

Scope of supply

To be confirmed for the selected model. Standard supply, grips, fixtures and options are listed separately for the selected configuration.